BIOLINE Manual for installation and use
OPTIKON 2000
PARAMETER SPECIFICATIONS
TECHNICAL SPECIFICATIONS
Probe: piezoelectric with fixing light, which can be
assembled on tonometric slide
Frequency: ..................................approx. 10 MHz
Clinical accuracy: ........................ 0.1 mm.
Resolution: .................................. 0.06 mm.
Field of measurement:................. 18-36 mm.
Screen:........................................ Liquid crystals 240x128 dots.
Speed of presentation: ................ 20 frames per second.
Printer:.........................................40 columns - 320 dots, thermal.
Modes of measurement:.............. Automatic and manual.
Measurements in the memory: .... up to 5 with indication of average and
standard deviation
Echo amplification: ...................... Manual in 7 levels or Self-regulation
Calculation formula:..................... SRKII , Haigis, Holladay, Gernet, SRK/T
Lens memory:.............................. Up to 15 lenses, with relating storable parameters
Acoustic signalling:...................... Sound or voice synthesis, selectable
Outputs
For computer: RS232, with 9 pin connector.
For the optional footswitch, with 2 way jack
Acoustic Output Parameters according to IEC/EN/DIN 61157 standard
See page 5-4
Code 141002EN 5-3 26.10.05 Rev.0