32
16. Technical data
OSM2 SI-923 SI-915
Supply voltage: 220 – 240 V 100 – 130 V
Rated current: 0.1 – 0.8 A 0.2 – 1.6 A
Permitted voltage fluctuation: ± 10 % ± 10 %
Frequency: 50 – 60 Hz 50 – 60 Hz
Max. mechanical output power: 70 W 70 W
Max. torque on the motor: 5.5 Ncm 5.5 Ncm
Max. coolant in ml (with hose ø 6/4): approx. 100 ml/min approx. 100 ml/min
Fuses: 2x 250 V – 1.25A slow-blow 2x 250 V – 1.25A slow-blow
Operating mode: S3 (4min/10min) S3 (4min/10min)
Dimensions in mm (WxDxH): 252 x 254 x 90 252 x 254 x 90
Weight in kg: 3.30 3.30
The foot control is water-tight according to IPX8, 1m depth of immersion, 1 hour
(water-tight according to IEC 60529)
Equipment of Protective Class II (type of protection against electric shock)
Type BF appliance (not suitable for intracardiac application)
Motor REF 04720024 and foot control REF 04785700 conform to Class AP
according to IEC 60601-1 / UL 60601-1 in danger zone M.
Classification according to § 5 of the General Specifications for the Safety of Medical Electrical Equipment
according to IEC 60601-1 / UL 60601-1