Disinfection, Cleaning, Sterilization
Motor with cable
Do not twist and kink the motor cable! Do not coil it too tightly! This instruction is not necessary when a sterilization cassette
is utilized (optional accessory).
Pre-disinfection
– If heavily soiled: Clean first with disinfectant cloths.
Manual cleaning
– Rinse under demineralized water (< 38 °C) with the aid of a brush (not recommended in U.K.)
– Then remove any liquid residues (absorbent cloth, blow dry with compressed air)
Disinfection
– Disinfection with surface disinfectants, wiping disinfection is recommended.
– Use only surface disinfectants which do not contain chlorine and which are certified by officially recognized institutes.
For USA: Use EPA registered surface disinfectants.
– Note the dosage and the action time of the surface disinfectants used!
Before disinfection and cleaning
push the motor protective cover
onto the motor attachment!