EasyManua.ls Logo

oticon Intent 2 miniRITE - Page 59

oticon Intent 2 miniRITE
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
59
Radiated emissions requirements for a
CISPR11 Group 1 Class B device as stated
in table 2 of IEC60601-1-2.
Radiated emission for communication
devices operating in the ISM 2.4GHz
band as stated in 47CFR Part 15,
subpartC, RSS-247 and EN300328
(only for products with Bluetooth).
Radiated emission for near field
magnetic induction communication
operating at 3.84MHz as stated in
47CFR Part 15, subpartC, RSS-210 and
EN300 330.
RF radiated immunity at a field level of
10V/m between 80MHz and 2.7GHz
as stated in table 4 of IEC60601-1-2
and, 3V/m between 2.7 and 6.0GHz as
stated in CISPR32.
RF radiated immunity at a field level of
up to 28V/m for selected RF wireless
communication bands between 380MHz
and 5.8GHz as stated in table 9 of
IEC60601-1-2.
RF radiated immunity at a field level
of up to 65A/m for selected proximity
magnetic fields at 30kHz, 134.2kHz
and 13.56MHz as stated in table 11 of
IEC60601-1-2 AM1.
RF radiated immunity field levels at
selected bands stated in IEC60118-13
between 25V/m and 60V/m and, for
proximity-fields from communication
devices held to the ear, stated in
ANSIC63.19.

Related product manuals