Coupler adapter
The Accessory Box provides a range of adapters for easy positioning of different types of hearing instruments.
2 cc coupler cavity
During tests in the test chamber, the hearing instrument is connected to a 2cc coupler cavity manufactured in accordance
with the ANSI standard.
Coupler microphone
The coupler microphone is located in a coupler bottom piece which must be attached to the 2cc coupler cavity.
You can use the coupler microphone either directly in AURICALHIT or in the Accessory Box.
In AURICALHIT
A. BTE testing - Low coupler position
B. ITE, RIE, thin-tube testing - High coupler position
In the Accessory Box
Connect the mini-jack cable from the Accessory Box to the mini-jack socket under AURICALHIT, and insert the coupler
microphone in the microphone socket in the Accessory Box.
A. Wireless hearing instrument testing
Otometrics - AURICALHIT
7
4 The test chamber