EasyManua.ls Logo

Owon XDM1041 - Page 21

Owon XDM1041
45 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1.Safety Information
DC Current
4. Set the range.
Press the Range softkey to set the range. Auto range automatically selects the
range for the measurement based on the input.
Note:
The multimeter uses two kinds of fuses for current protection: A 10A fuse is
built into the 10A input terminal for overcurrent protection, and a 600mA
fuse is built into the uAmA input terminal for overcurrent protection.
10% over range for all ranges except 10 A range.
If the reading exceeds 10.5 A in 10 A range, "overload" will be displayed.
5. Set the measurement speed.
Press the Speed softkey to switch between Low, Mid and High. See page 14,
Measurement Speed .
6. Set the relative value.
Press the Rel softkey to turn on or o the relative operation. For relative
operation, the multimeter subtracts the pre-specied value of REL operation
from the actual measurement result and displays the result. See page 29,
Relative Value.
Measuring AC Current
This section describes how to congure AC current measurements.
Operating Steps:
1. Enable the ACI measurement.
Press on the front panel, press it again to enter ACI measurement
mode.
17
All manuals and user guides at all-guides.com

Other manuals for Owon XDM1041

Related product manuals