15. After the last measurement has been taken, enter the thickness of the
shim using the numerical keys. Once a numerical key has been pressed,
the probe can be lifted from the shim.
16. Check the display. If incorrect, press the CLEAR key and then re-enter
the number. When finished, lift the probe at least the probe at least 3
inches (76 mm) away from the part and press the ENTER key.
17. The 233 gage is now calibrated and ready for measurement.
MEMORY MANAGEMENT
Memory Structure
The 233 has 99 memory locations. Each location stores a calibration, statistics
and readings. Locations 1 to 50 are allocated for Magnetic applications and
location’s 51 to 99 for Eddy Current applications. The memory location is shown
in the upper left of the display. “FE” shows in the lower left of the display for
Magnetic applications and “NON-FE” shows for Eddy Current applications.
When the gage is turned on (and the probe has not been changed), it selects the
last memory location that was used to take measurements and displays the last
reading for the application last used to take a measurement.
How to Select A Memory Location
A memory location must be made active before it can be used to store
information. To activate a memory location, select it by using Method 1. To
select memory locations that already are active, use either of the following
methods.
Method 1
1. Press the SEL key. The word SEL will appear on the display.
2. Enter the desired memory location (1-99) using the numerical keys. The
location number will appear in the upper left of the display.
3. Press the ENTER key.
4. The unit will beep twice if this memory location was just activated and
0.000 will show in the main display, otherwise the unit will beep once and
the last reading taken will show in the main display.
Method 2
1. Press the SEL key. The word SEL will appear on the display.
2. Press the SEL key again, causing the display to scroll through the active
memory locations. The location number will appear in the upper left of the
display.
3. When the desired memory location is displayed, press the ENTER key.
4. The last reading taken will show in the main display.
How to Clear a Memory Location
When a memory location is cleared it is deactivated and all statistics,
measurements, and parameters are deleted. If memory location 1, and /or
location 51 are cleared, all memory locations that are assigned to that universal
calibration will also be cleared. A memory location can be cleared using one of
the following methods.