How to Set Limits
1. Press the LIMIT key. H-L will show in the main display.
2. To enter the high limit, press the HIGH key. Or to enter the low limit,
press the LOW key.
3. The default (See Setup – Applications for the range for each application)
or any previously entered limit will be displayed.
4. To accept press ENTER key, otherwise enter the desired limit using the
numerical keys and press the ENTER key.
How to Remove Limits
1. Press the LIMIT key. H-L will show in the main display.
2. To remove the high limit, press the HIGH key. Or to remove the low limit,
press the LOW key.
3. For high limit enter default high limit shown in the table above for Setup -
Applications, for low limit enter 0.000 mil/ 0.000 um and press the ENTER
key.
DATE AND TIME STAMP
The 233 has a Date and Time Stamp feature. This feature is supplied in one of
the two options described in Memory Capacity. The date and time information
will be included when statistics and readings are printed, or uploaded to a PC.
Memory Capacity
The 233 can be ordered with the Date and Time Stamp Every Reading feature
installed. The user can then turn it on or off. See Printing the Date and Time
below. With this feature installed the date and time is stored with each reading.
The maximum number of readings saved is 8900.
The default installation is Date and Time Stamp First and Last Reading. With
this option the date and time is stored only for the first and last reading taken at a
memory location. With this option, 26000 readings can be saved.
Setting the Date and Time
Before taking measurements with the gage, the correct time and date should be
entered. This information will be stored with the readings and appear on the
printouts or uploads.
1. Press in sequence the “٭” “1” “0” keys, and then press the ENTER key.
2. The word “cL” will show in the upper left of the display and a time in the
main display. The time displays and prints in a 24-hour format. Example:
1:18 P.M. will display as 1318.
3. Enter the correct time in a 24-hour format. Press the ENTER key if
entering three digits and the display will advance to the next setting.
4. The word “dA” will now show in the upper left of the display and a date in
the main display. Key in the value for the month and day in the format of
MMDD and press the ENTER key.
Oxford Instruments Contact Measurement Page 15 of 24
233 USER GUIDE
The information in this document is private and confidential and must not be copied.
Oxford Instruments plc