Problem 7: The gage locks up and no other troubleshooting techniques correct
the problem.
Solutions 7: Use this procedure as a last resort. Try all other solutions first.
Clear all of the gage settings by pressing in sequence the٭” “9” “1” “9” “1” keys.
This clears all calibrations, statistics, and probe tuning information from every
memory location. After this procedure is used, retune all probes and re-calibrate
the gage.