-12-
6.8.3 1 Point Calibration
Steps :
1. Select Calibration from the applications.
2. Select Enter Password from the sub menu by default, press
“0000” if you have not set your own password.
3. Select 1 Point Cal. from the sub menu.
4. Select Expected Volume1. By default, 10% of the nominal
volume will be on the display, you can change this volume
according to your requirement by using up / down arrow key.
5. Enter Delivered Volume1.
6. Select “Yes” to save the changes. It will show Calibration
done on the display.
7. Select “No” to exit.
Steps :
1. Select Calibration from the applications.
2. Select Enter Password from the sub menu. By default, press
“0000” if you have not set your own password.
3. Select Restore Fact. Cal.
4. Select Yes by pressing up / down arrow key to restore the
factory calibration. “Factory Calibration Restored” screen will
appear for few seconds and you will directed to the
calibration menu.
5. Select No by pressing up / down key to keep the user
settings.
Special Note : Whenever the user makes any changes in the calibration,
the Applications will show “Calibration User”. Once you restore the
factory calibration, Applications will show “Calibration Fact.”.
6.8.4 Restore Factory Calibration
This feature will allow you to restore all the factory generated data of
the product. So, all the user changes will be nullified and factory
calibration will be restored.
Steps :
6.8.5 Last Calibration Date
Here you can keep a track of your last calibration date. Whenever you
calibrate your pipette, you can enter the date of the calibration.
1. Select Calibration from the applications.
2. Select Enter Password from the sub menu. By default, press
“0000” if you have not set your own password.
3. Select Last Cal. Date by using up / down arrow key.
4. Enter the date in DD/MM/YY format and you will be
directed to the calibration menu.