9.1 Single Channel Pipette Technical Specification
-15-
Cat. No.
OBE-10
Test Volume
(µl)
Inaccuracy
(±) %
Imprecision
(±) %
0.5-10 µl
Volume
Range (µl)
1
5
10
2.5
2
1
1.8
1
0.6
OBE-100
5-100 µl
10
50
100
3
1
0.8
1
0.5
0.2
OBE-200
10-200 µl
20
100
200
2.5
0.7
0.6
0.7
0.3
0.2
OBE-1000
50-1000 µl
100
500
1000
3
1
0.6
0.6
0.4
0.2
OBE-5000
250-5000 µl
500
2500
5000
2.4
1.2
0.6
0.6
0.25
0.15
Color
Cat. No.
Test Volume
(µl)
Inaccuracy
(±) %
Imprecision
(±) %
0.5-10 µl
Volume
Range (µl)
1
5
10
8
4
2
5
2
1
OBE-8-20
OBE-12-20
2-20 µl
2
10
20
7
3
2
3
2
1.6
OBE-8-100
OBE-12-100
5-100 µl
10
50
100
3
1
0.8
2
0.8
0.3
OBE-8-200
OBE-12-200
10-200 µl
20
100
200
5
1
0.7
1.4
0.4
0.25
Color
9.2 Multi Channel (8-12) Pipette Technical Specification
OBE-8-10
OBE-12-10
OBE-8-300
OBE-12-300
15-300 µl
30
150
300
3
1
0.6
1
0.5
0.3