-8-
Steps :
1. Select Diluting from applications.
2. Select Diluting + Mixing.
3. Select No. of Dilutants (Max. 2 dilutants can be selected).
4. Select 1st volume by using up / down arrow key.
5. Select 2nd volume by using up / down arrow key.
6. Select Mixing Cycle by using up / down arrow key. Max. 10
cycles can be selected.
7. Aspirate 1st volume by pressing trigger key.
8. Aspirate Air by pressing trigger key.
9. Aspirate 2nd volume by pressing trigger key.
10. Dispense total volume in vessel by using trigger key.
11. Mixing will automatically start once you dispense the liquid
into the vessel.
12. You can stop mixing if require by using the ESC key.
13. Press trigger to start the mixing again / Press the ESC key to
completely stop the mixing in between the process. It will discard
the excess liquid remaining in the tip.
6.5.2 Diluting + Mixing
6.6 Mixing
You can mix the liquids using this mode. There are three modes under
mixing.
a) Mixing . b) Pipetting + Mixing. c) Diluting + Mixing.
6.6.1 Mixing
Steps :
1. Select Mixing from the applications.
2. Select volume by using the up/down arrow key.
3. Select Mixing Cycles by using up/down arrow key. Max. 10
cycles can be selected.
4. Aspirate the volume.
5. Mixing will automatically start .
6. You can stop mixing by using the ESC key.
7. Press trigger to start the mixing again / Press the ESC key to
completely stop the mixing in between the process. It will discard
the excess liquid remaining in the tip.