Lumiso
Pooltest
A Palintest Product
LMP001
Blank
530nm
575
nm
12:28:49
18-01-2021
9
Continue, by inserting
Standard B, pressing read,
and comparing the result
with certificate.
Then repeat the entire process
(blank, read A, read B) from
step 2 onwards for the second
wavelength.
Should any standards read outside the values shown
on your certificate, the most likely cause is contamination
on the optical windows in the cell holder. Please clean
the cell holder carefully and remeasure the standards.
(See ’Keeping the Optics Clean’)
If this does not resolve the issue, please contact your local
Palintest branch or distributor.