Instructions for Use
Allegro™ 200 L and 500 L Plastic Totes, Lids
and Trolleys
For use with Allegro 3D biocontainer systems and bag systems
from other manufacturers (100L, 200L and 500 L)
(Part numbers LGRPTTE200L, LGRPTTEL200L, LGRPTRL200L,
LGRPTTE500L, LGRPTTEL500L and LGRPTRL500L)

Questions and Answers

  • S
    sherry37Aug 9, 2025
    Why is my Pall Laboratory Equipment system not filling or draining properly?
    • J
      Jessica HarrisAug 9, 2025
      The fluid supply to your Pall Laboratory Equipment might be restricted or closed. Check the fluid supply, ensuring that all clamps on the appropriate line are open and the tubing isn't pinched or twisted. Also, make sure all connections on the appropriate line are fully activated.
  • M
    Mitchell LawrenceAug 14, 2025
    What to do if the Allegro biocontainer system does not open correctly during fill in Pall Laboratory Equipment?
    • G
      garyadamsAug 15, 2025
      The Allegro biocontainer might not open correctly during filling because it was unfolded prior to installation. These systems are designed to self-unfold and self-position, so ensure the side panels are folded over the top face of the system when installed. Another reason might be that the Allegro system is installed with the folded sections at the front and back of the tote instead of at the sides. Reinstall the system, checking that the corners of the biocontainer are aligned with each of the tote's corners.
  • S
    Stephanie AguilarAug 21, 2025
    What to do if there are excessive creasing and folds during filling of Pall Laboratory Equipment system?
    • R
      Robin RodriguezAug 22, 2025
      Excessive creasing and folds during filling may occur if the selected system is too large for the tote being used. Make sure the correct single-use system is selected and properly installed in the chosen tote.
  • E
    Emily RamirezAug 27, 2025
    What to do if the Pall Laboratory Equipment system does not form to the sides of the system?
    • G
      Gina JonesAug 27, 2025
      If the system doesn't form to the sides, it might be because the selected system is too small for the tote being used. Double-check that the correct single-use system is selected and properly installed in the chosen tote.
  • J
    Jonathan WilliamsSep 2, 2025
    How to fix Pall Laboratory Equipment if the system does not align with the corners of the tote?
    • J
      Jacob PerezSep 2, 2025
      If the locating port bracket is securely in position, yet the system doesn't align with the tote's corners, it might be due to the locating port bracket being installed in reverse. Confirm that a Pall Allegro biocontainer system is being used. If so, align the port bracket as shown in Figure 34. If you're using another manufacturer’s biocontainer, reverse the bracket as per Figure 35.

Summary

Warnings

Receipt of Equipment - Allegro Plastic Totes and Allegro Single-use Systems

Installation and Operation

System Preparation, Installation and Use of Allegro Single-use Biocontainer Systems in an Allegro Plastic Tote

Detailed steps for preparing, installing, and using Allegro single-use biocontainer systems within Allegro plastic totes.

Installation of a Single-use Biocontainer System in the 200 L tote

Specific instructions for installing a biocontainer system into the 200 L Allegro plastic tote.

Installation of a Single-use Biocontainer System in the 500 L tote

Specific instructions for installing a biocontainer system into the 500 L Allegro plastic tote.

Filling an Allegro 3 D Biocontainer System

Step-by-step instructions for filling Allegro 3D biocontainer systems, including safety warnings.

Maintenance and Spare Parts

Troubleshooting Guide and Useful Tips