EasyManua.ls Logo

Panasonic BT-H1700P

Panasonic BT-H1700P
121 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
46
CONTRAST ADJUSTMENT (SDTV)
Measuring Instruments Signal generator (Crosshatch signal)
Oscilloscope
Card (Slot) Component/RGB Input Card (Slot 1)
Test Points TP-47G [CRT SOCKET PWB]
TP-GND [CRT SOCKET PWB]
Adjustment Points S*01 (4:3 Overscan Contrast), SM** (4:3 Underscan Contrast),
SN** (16:9 Overscan/Underscan Contrast) [Service Menu]
Notes: Perform the following adjustments after completing the
1080/60i signal Contrast adjustment.
Set the CONTRAST data in the Setup Menu to "00".
The SO value (see the description of the HDTV Contrast
Adjustment) becomes the reference value for the follow-
ing adjustments. When this data is changed, it is required
to re-adjust the data of all of the adjustment signals (HDTV,
SDTV and NTSC/PAL).
When re-adjusting the 1080/60i signal, use the SI.
When overscan data of a signal is changed, it is required
to re-adjust the underscan data of the same adjustment
signal.
When the 4:3 data of a signal is changed, it is required to
re-adjust the 16:9 data of the same adjustment signal.
1. Apply the 480/60i crosshatch signal to INPUT A (Terminal Y on the
Component/RGB Input Card).
2. Set the CONTRAST potentiometer on the front panel to the center
click position.
3. Connect the oscilloscope across TP-47G and TP-GND.
4. Adjust SE01 in the Service Menu to set the voltage amplitude <A>
in the figure on the right to 30 V ± 2 V.
5. Set the ASPECT button on the front panel to ON so that the scan-
ning size is 16:9.
6. Adjust SN25 to set the voltage amplitude <A> to 30 V ± 2 V.
7. Set the ASPECT button on the front panel to OFF and the UNDER
SCAN button to ON to set the scanning size to underscanning for
the 4:3 ratio.
8. Adjust SM13 to set the voltage amplitude <A> to 30 V ± 2 V.
9. Set the ASPECT button on the front panel to ON to set the scanning
size to underscanning for the 16:9 ratio.
10. Adjust SN28 to set the voltage amplitude <A> to 30 V ± 2 V.
11. Set the ASPECT and UNDER SCAN buttons on the front panel to
OFF.
12. Vary the adjustment signal and adjustment data, and re-perform
adjustments in steps 1 to 11 above (see Table 2).
1 Vertical interval
(V. sync)
<A>
30V ± 2V
Adjustment
Scanning Size
Aspect Ratio
Signal
4 : 3 16 : 9
480/60i
Overscan SE01 SN25
Underscan SM13 SN28
576/50i
Overscan SF01 SN31
Underscan SM16 SN34
480/60p
Overscan SG01 SN37
Underscan SM19 SN40
Table 2

Table of Contents

Related product manuals