31
TAKING
A
READING
(CONTINUED)
IRREGULAR
PULSE
DETECTION
FUNCTION
If
the
heart
rate
measured
exceeds
the
average
value
for
more
than
+25%
during
measurement,
the
symbol
of
"Irregular
Pulse"
will
be
displayed
to
remind
that
atttention
is
needed
after
measurement.
After
measurement,
the
symbol
of
"Irregular
Pulse"
will
be
displayed
after
6
seconds
of
blinking.
*
When
the
symbol
of
"Irregular
Pulse"
is
displayed,
please
measure
again.
«lf
the
heart
rate
changes
too
much
during
measurement,
it
may
affect
the
accuracy
of
measured
value.
Flashes.
Is
lit
*
Blood
pressure
values
are
displayed
alternately
in
mmHg
and
kPa
units.
e
When
you
save
and
select
the
value
measured
with
irregular
pulse,
the
symbol
of
"Irregular
Pulse"
will
be
displayed
as
well.
This
symbol
is
not
designed
to
indicate
irregular
pulse.
e
This
is
a
reference
function
intended
for
ensuring
the
accuracy
of
measured
value.
In
order
to
ensure
the
accuracy
of
measurement,
please
measure
for
2-3
times
under
static
status
even
if
the
symbol
of
"Irregular
Pulse"
is
not
displayed.(see.p.27)
e
If
the
symbol
appears
constantly,
you
should
consult
a
doctor
regarding
your
health
condition.
Please
do
not
estimate
the
result
and
receive
treatment
by
yourself;
Doctor's
advice
must
be
followed.
Reading
the
blood
pressure
display
Whether
the
systolic
and
diastolic
blood
pressure
values
are
in
the
high
blood
pressure
region
or
normal
blood
pressure
region
are
judged
based
on
the
WHO/ISH(Word
Health
Organization/International
Society
of
Hypertension)blood
pressure
classifications,and
the
blood
pressure
value
blinks
if
required.
WHO/ISH
blood
pressure
classifications%
|
90
Diastolic
(mmHg)
If
the
measured
blood
pressure
value
is
in
the
high
blood
pressure
region,the
values
will
blink
for
approximately
6
seconds.
+
Systolic
blood
pressure:18.7kPa
(140mmHg)
or
higher
*
Diastolic
blood
pressure:12.0kPa
(90mmHg)
or
higher
(Only
immediately
after
measurement)
(1kPa=7.
5mmHg)
If
measurements
could
not
be
taken
properly
85
The
remeasure
mark
"E"
will
displayed.
When
taking
successive
measurements,
press
the
开始
/结束
button
to
switch
off
the
blood
pressure
monitor
once
and
stay
quiet
for
4
to
5
minutes,and
then
take
the
next
measurement.
Advice:
Often
with
the
second
measurement,there
is
less
tension
and
you
feel
more
relaxed,so
the
blood
pressure
value
is
lower.
For
people
who
easily
get
tense,measuring
twice
each
time
is
very
effective.
When
two
measurements
are
taken,normally
the
smaller
measurement
is
stored.When
taking
two
measurements,always
stay
quiet
for
4
to
5
minutes
after
the
first
measurement
before
taking
the
second
measuremnt.
*4
1999
World
Health
Organization-International
Society
of
Hypertension
Guidelines
for
the
Management
of
Hypertension
32