4 CLEANING AND DISINFECTION
DANGER
Danger of electrocution
Liquids can conduct electricity, thereby posing a risk of elec-
tric shock.
Always switch the compressor off and disconnect the
power plug from the mains socket before starting to clean
it.
If the inhalation system is used by only one patient, follow the
instructions for cleaning and disinfecting in the section "Single
patient use" [see:Single patient use, page28].
If the inhalation system is used by several patients, follow the
instructions for cleaning and disinfecting in the section "Multiple
patient use" [see:Multiple patient use, page33].
4.1 Single patient use
Cleaning and disinfection cycles
Compressor housing Clean when visibly dirty
Nebuliser (without
connection tubing) and
mask
Clean immediately after every use
Disinfect at least once a day
Air filter Replace after 200operating hours
(approx.1year)
Nebuliser and mask
The mask can be cleaned, disinfected and dried together with
the nebuliser.
Preparation
MASK
Detach the mask from the nebuliser.
Detach the elastic band from the mask.
– 28 –