Cleaning the compressor housing
NOTE
Danger of device fault due to liquid penetration
If liquids get into the interior of the compressor, this may
cause a fault in the device.
• Never immerse the compressor in water.
• Never clean the compressor in running water.
• Never spray any liquids onto the compressor or the power
cord.
• If liquid does get into the compressor, it must not be used
under any circumstances. Before starting the compressor
again, contact the TechnicalService department or the
ServiceCenter of PARIGmbH.
Clean the compressor housing as follows:
• Wipe the outer surface of the
housing with a clean, damp
cloth.
Air filter
Take the air filter out of the compressor and inspect it at regu-
lar intervals (after every 10applications). If it is discoloured
brown or grey, or if it is damp or clogged, replace it.
The filter must be replaced after 200operating hours (ap-
prox.1year) at the latest [see:Replacing the air filter,
page34].
The air filter cannot be cleaned and then reused!
– 32 –