EN | User Manual Maintenance and repair of the dryer
BSP-MT 10-95/D3 37
Maintenance work to be completed every 12 months
Check air quality
Especially for critical applications, we recommend to annually check the quality of the pro-
cessed medical breathing air (this check should be carried out by a quali ed authority).
Renew muffl ers
The dryer is equipped with a muf er. If the muf er becomes blocked, a dam pressure is gen-
erated which in extreme cases may cause the muf er to burst.
Hazard caused by blocked muffl er!
Blocked muffl ers can cause a dangerous overpressure to build up which may
cause the muffl ers to burst. Flying fragments may cause personal injury and
damage to property.
Therefore, the muffl ers must be replaced every 12 months and after each
change of desiccant.
Warning against sudden air ejection!
During expansion the pressure is released suddenly through the muffl er:
◊ A loud cracking noise occurs which can injure your hearing.
◊ Particles carried in the air fl ow act like bullets and can injure your eyes or
skin.
Always wear eye and ear protection, therefore, when you are in the vicinity of
the dryer!
Depressurise the dryer and shut it down
(see page 30 ).
Unscrew muf er as shown in the opposite
gure.
Replace muf er and secure it.
Restart dryer (see page 31).
Undo muf er
Replace the element in the fi ne fi lter muffl er
Depressurise the dryer and shut it down
(see page 30 ).
Undo knurled screw on the lid cap and
remove cap.
Unscrew old lter element. Arrows on the
element bottom mark the direction of rotation.
Insert new lter element and screw on
tightly.
Renew muf er element
Re-engage lid cap at the top section of the housing and x in position by means of the
knurled screw.
Restart dryer (see page 31 ).