MAXIGAS
MX104 - MX120
Original Language
User Guide
EN

Questions and Answers

  • D
    Daniel JonesAug 7, 2025
    Why is there no gas outlet pressure in my Parker Portable Generator?
    • E
      Erica AllenAug 7, 2025
      If your Parker Portable Generator shows no or low gas outlet pressure, it could be due to several reasons. First, the generator might be overdue for service, in which case you should service it. Second, there might be an internal or external gas leak, which you should check and rectify. Finally, the inlet pressure might be low; ensure the pressure meets the required specifications.
  • A
    Amy CantuAug 12, 2025
    How to fix a Parker Portable Generator with no LCD and status indicators?
    • R
      Russell AndrewsAug 12, 2025
      If the LCD and status indicators on your Parker Portable Generator are not illuminated despite being connected to power, the issue might stem from a blown fuse. In this case, replace the fuse. Alternatively, the ribbon cable may be disconnected; try reconnecting it. Finally, ensure that the power is properly connected; if not, reconnect the power.
  • A
    alvintranAug 17, 2025
    What to do if Parker Portable Generator is making excessive noise?
    • B
      berrysarahAug 17, 2025
      If your Parker Portable Generator is producing excessive noise or vibration, it could be due to a loose or defective silencer. Check and rectify the silencer. It could also be caused by solenoid valve wear or a loose coil. Check the solenoid valve and replace it if necessary.
  • N
    Nicole JordanAug 22, 2025
    What causes high oxygen concentration in Parker Portable Generator?
    • J
      Justin RobinsonAug 22, 2025
      If your Parker Portable Generator shows a high oxygen concentration, it could be caused by a defective oxygen cell, which should be replaced. Alternatively, there might be a leak in the system piping, which needs to be checked and rectified.
  • A
    Alexander DavisAug 27, 2025
    What to do if Parker Portable Generator has high outlet pressure?
    • T
      Tracy StoneAug 27, 2025
      If your Parker Portable Generator has a high outlet pressure, the outlet regulator might be defective. Try resetting or replacing it.
  • M
    marysullivanSep 1, 2025
    Why does my Parker Portable Generator have low inlet pressure?
    • Z
      Zachary JacksonSep 1, 2025
      Low inlet pressure in your Parker Portable Generator can be caused by a few things. The compressor or ring main pressure might be low; check and rectify this. Also, the inlet valve might not be open; check and rectify this as well. If these steps don't solve the problem, refer to the pre-treatment manual for defects on the pre-treatment package.

Summary

Safety Information

Description

Technical Specification

Provides detailed technical specifications for the MAXIGAS nitrogen generator, including dimensions and parameters.

Installation

Electrical Installation

Details field wiring requirements, cable gland usage, and electrical connections for generator and dryer supplies.

Generator Supply

Specifies requirements for the main electrical supply connection, including switch/circuit breaker ratings.

Operating the Generator

Starting the generator

Step-by-step instructions for safely starting the generator, including initialisation and pressure checks.

Stopping the generator

Procedures for safely shutting down the generator, including closing the outlet valve and initiating shutdown.

Fault Log

Accessing and understanding the fault log, including fault codes, probable causes, and their meanings.

Customer Settings

Details how to alter generator parameters via the customer settings menu, including O2 alarm and economy settings.

Oxygen Analyser Calibration

Procedure for calibrating the O2 analyser using calibrated gas or an independent analyser for accuracy.

Entering the calibrated level

Steps for entering the calibrated level into the controller after performing the O2 analyser calibration.

Servicing

Troubleshooting

Parker MAXIGAS MX106 Specifications

General IconGeneral
BrandParker
ModelMAXIGAS MX106