Questions and Answers

  • R
    Roberto KingAug 3, 2025
    Why is there no gas outlet pressure in my Parker Portable Generator?
    • M
      Mary WalkerAug 3, 2025
      If your Parker Portable Generator shows no or low gas outlet pressure, it could be due to several reasons. First, the generator might be overdue for service, in which case you should service it. Second, there might be an internal or external gas leak, which you should check and rectify. Finally, the inlet pressure might be low; ensure the pressure meets the required specifications.
  • S
    Sherry CarterAug 7, 2025
    How to fix a Parker Portable Generator with no LCD and status indicators?
    • P
      Peter GreenAug 7, 2025
      If the LCD and status indicators on your Parker Portable Generator are not illuminated despite being connected to power, the issue might stem from a blown fuse. In this case, replace the fuse. Alternatively, the ribbon cable may be disconnected; try reconnecting it. Finally, ensure that the power is properly connected; if not, reconnect the power.
  • L
    lindsay65Aug 12, 2025
    What to do if Parker Portable Generator is making excessive noise?
    • R
      rgarciaAug 13, 2025
      If your Parker Portable Generator is producing excessive noise or vibration, it could be due to a loose or defective silencer. Check and rectify the silencer. It could also be caused by solenoid valve wear or a loose coil. Check the solenoid valve and replace it if necessary.
  • C
    Christina ThompsonAug 17, 2025
    What causes high oxygen concentration in Parker Portable Generator?
    • M
      markromeroAug 17, 2025
      If your Parker Portable Generator shows a high oxygen concentration, it could be caused by a defective oxygen cell, which should be replaced. Alternatively, there might be a leak in the system piping, which needs to be checked and rectified.
  • C
    Charles WilliamsonAug 22, 2025
    What to do if Parker Portable Generator has high outlet pressure?
    • K
      Kendra AllenAug 22, 2025
      If your Parker Portable Generator has a high outlet pressure, the outlet regulator might be defective. Try resetting or replacing it.
  • B
    Barry ChambersAug 27, 2025
    Why does my Parker Portable Generator have low inlet pressure?
    • J
      jesus58Aug 27, 2025
      Low inlet pressure in your Parker Portable Generator can be caused by a few things. The compressor or ring main pressure might be low; check and rectify this. Also, the inlet valve might not be open; check and rectify this as well. If these steps don't solve the problem, refer to the pre-treatment manual for defects on the pre-treatment package.

Summary

Safety Information

Product Description

Technical Specifications

Detailed technical specifications including parameters, units, and model variations for the generator.

Installation Guide

Electrical Installation

Instructions for field wiring and electrical connections to the generator.

Generator Operation

Starting the Generator

Step-by-step procedure for starting the MAXIGAS nitrogen generator.

Stopping the Generator

Procedure for safely shutting down the MAXIGAS nitrogen generator.

Menu Interface Navigation

How to navigate through the generator's menu system and access various functions.

Customer Settings Configuration

Details on how to access and modify user-configurable parameters in the settings menu.

Oxygen Analyser Calibration

Procedure for calibrating the generator's oxygen analyser.

Entering Calibrated Level

Steps for inputting the calibration gas or measured value into the analyser.

Servicing and Maintenance

Service Procedures

Detailed steps for performing common maintenance tasks on the generator.

Troubleshooting Guide

Parker MAXIGAS MX104 Specifications

General IconGeneral
Fuel TypeGasoline
Engine Type4-Stroke OHV
Starting SystemElectric Start
Fuel Tank Capacity6.6 gal
AC Output120/240V