24 - EN
5 Preventative Maintenance
5.1 Cleaning
Clean the equipment with a damp cloth only and avoid excessive moisture around any electrical sockets. If required you may use a mild
detergent, however do not use abrasives or solvents as they may damage the warning labels on the equipment.
5.2 Maintenance Schedule
Key:
Description Of Service Required
Service Recommended Every:
1
1.The service operations should be performed at the hours run or fixed time intervals specified (whichever occurs first)
Component Operation
Daily
3 Months
(2000 Hrs.)
6 Months
(4000 Hrs.)
12 Months
(8000 Hrs.)
24 Months
(16000 Hrs.)
36 Months
(24000 Hrs.)
60 Months
(40000 Hrs.)
Generator Check the status indicators located on the front panel.
System Check the inlet air quality.
Generator Check for air leaks
Generator Check the pressure gauges during purging for excessive back pressure.
Generator Check the condition of the electrical supply cables and conduits.
Generator Check oxygen sensor(s) and calibrate if necessary
Generator Check for cyclic operation
Filtration
Replace Exhaust silencer and filter element(s)
Recommended Service A
Generator
Replace Oxygen sensor(s)
Recommended Service B
Generator
Replace control valves
Recommended Service C
Generator
Replace cylinder and solenoid valves
Recommended Service D
Check
(Operator)
Essential Procedure
(Service Personnel only)
Essential Procedure
(Responsible Body or Service Personnel only)