20
SENSOR PORT: P1
CUSTOM CAL....
SENSOR PORT: P1
SENSOR ZERO: NO
YES & ENTER TO ZERO
CAL PRESS: 60.00 psi
CUSTOM CAL SENSOR
PRESS ENTER TO CAL
CALL IN PROGRESS
Please Wait: 1...2...
CUSTOM CAL FACTORS:
CF: 0.511 PZ: 0.0
CALIBRATION ERROR!
OUT OF RANGE
SENSOR RESET: NO
YES & ENTER TO RESET
See 3.1
See 3.2
See 3.3
See 3.4
See 3.5
See 3.6
3.0 Custom Calibration Menu
Press “Sensor” button. The
following will be displayed:
Enter
Enter
Enter
Enter
Enter
See 3.7
Press Enter
Use up / down arrows to
choose port, press “Enter”, and
calibration data will reload.