1.2 PC Terminal Program Settings
SciPres
®
pressure monitor to PC: PC communications via the Serial Port labeled “Printer / PC”
requires a SciPres
®
RS-232 Cable.
NOTE: When you are not using the SciPres
®
with a printer, this allows process data to be “dumped”
into your PC for archiving. The list of settings below must match those in serial section of the
SciPres
®
pressure sensor main menu.
The following terminal setting procedure is intended for PCs with a Windows 95/98/XP software
installation (for any other operating system, download and use a terminal or Telenet program and set
it up in a similar manner). Press the “START” button in the lower left corner of your screen, select
All Programs” then select and open “Accessories”, select “HyperTerminal”, double-click at the
HyperTerminal icon.
1. From the “Connection Description” screen, select an icon and enter a file name, i.e. SciPres
®
pressure sensor. Press “OK”.
2. From the “Phone Number” screen, select “Direct to Com 1” in the box labelled Connect Using: or
select any other available Com port. Press “OK”.
3. From the “Com 1 Property” screen, select the following parameters
Bits per Second: 9600
Data Bits: 8
Parity: None
Stop Bits: 1
Flow Control: None
Press “Ok”
4. Go to “File” and open “Properties”, from the “SciPres
®
Pressure Sensor Property” screen, select
the “Setting” screen, enter the following:
Terminal Keys: Select
Emulation: TTY
Backscroll Buffer: 500
While still in the “Setting” screen, press “ASCII Setup” key & select the following from the “ASCII
Setup” screen:
Send Line Ends with Line Feed: Select
Echo Typed Characters Locally: Select
Line Delay 10 msec
Character Delay: 10 msec
Append Line Feeds….....: No Selection
Force Incoming…............: No Selection
Wrap Lines That Exceed …..: Select
Press “OK” at the bottom of the “SciPres
®
Pressure Sensor Setup” screen.
Press “OK” at the bottom of the “SciPres
®
Pressure Sensor Property – Setting” screen.