EasyManuals Logo

Parr Instrument 6772 User Manual

Parr Instrument 6772
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #58 background imageLoading...
Page #58 background image
6772 Calorimetric Thermometer Instruction Manual
9-1
MAINTENANCE
Refer to Manual 204M, Operating Instructions for the 1341 Oxygen Bomb Calorimeter for
calorimeter maintenance information.
Refer to Manual 205M, Operating Instructions for the 1108 Oxygen Bomb for bomb
maintenance information.
CAUTION: Risk of Electrical Shock
Disconnect the electrical power before servicing or replacing any
components!
Fuses
The replacement of protective fuses for the 6722 Calorimetric Thermometer should be
performed by qualified personnel.
WARNING: For continued protection against possible hazard,
replace fuses with same type and rating of fuse.
Main line protective fuses. Rated Fast-Act 15 Amp, 250VAC (Parr number 139E23)
Fuse F1– Mains stirrer output fuse. Rated Fast-Act 1 Amp, 250VAC, time lag, 5 x 20mm
(Parr number 1641E)
This fuse is located on the A2140E I/O board located inside the instrument. To gain
access to the inside, first disconnect the power cord from the rear of the unit.
Remove the two screws at the upper rear of the unit that secure the front cover.
Pry up on the back lip of the cover and carefully hinge it forward. Make sure that
the cables attached to the front portion of the cover do not become stretched or
disconnected. Follow this procedure in reverse order to re-secure the front cover.
There is no routine maintenance recommended for the 6772 Calorimetric Thermometer.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Parr Instrument 6772 and is the answer not in the manual?

Parr Instrument 6772 Specifications

General IconGeneral
BrandParr Instrument
Model6772
CategoryThermometer
LanguageEnglish