Troubleshooting Flow Charts
32
PREVIOUS PAGE
Measure length signal of amplified output on connection board between test
point MP10 and test point MP13. (+5V)
NOTE: Negative signal at terminal X1:11 will be converted into positive signal
at MP13 (i.e.: input at terminal X1:11 = -0.5V and gnd).
Refer to system wiring diagram.
Correct?
No
Replace connection board and reset pressure channel.
Refer to system wiring diagram and main board
replacement Procedure.
Yes
END