Troubleshooting Flow Charts
37
PREVIOUS PAGE
Check power supply to console between console terminal 2(ground) and terminal 1 (+12v)
Refer to system wiring diagram.
Correct?
No
Faulty wiring in cable from central unit to console.
Replace cable.
Yes
Ensure that wires are properly connected between CU X1:31 and console terminal 4 and
between CU X1:32 and console terminal 3.
Refer to system wiring diagram.
Correct?
No
Replace main board and reset pressure channel.
Refer to system wiring diagram and transducer replacement
procedure.
Yes
END