7 Operation
7.1 Before first use
Fully charge the battery before the first use. The charging socket is located on the right
side above the sensor arm bracket.
Switch on the meter using the "On / Off" key and leave the device switched on for about
one hour for the sensor to acclimatise. Deactivate the automatic power-off function with
the APO key.
If you do not use the device for more than two months, repeat this procedure.
7.2 Before use
Check the meter for mechanical damage.
Only switch on the unit in an environment unpolluted by gas.
Carry out a gas test with the appropriate test gas.
Attention! When using a lighter, only move the sensor head very slowly towards
the gas source. A 100 % concentration exposure can destroy the sensor.
If the gas is not detected, the device is not ready for use.
After the gas test, wait until the display shows 0.