PCE Instruments PCE-P18 - Calculated Values; Selection of Controlling Values for Analog Outputs

PCE Instruments PCE-P18
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
16
AH = 2,1668
P
ws
RH
100 (T + 273,2)
gdzie:
T → measured temperature [
o
C]
RH → measured absolute humidity [%]
DP → dew-point temperature [
o
C]
P
ws
→ saturated water vapor pressure (water vapor pressure) [mbar]
AH → absolute humidity [g/m
3
]
DP =
T
n
m
- 1
log (P
ws
RH
10000 A
)
Table 2
The factors used to calculate the dew-point
T [
o
C] A m T
n
< 0
6.119866 7.926104 250.4138
0...50
6.1078 7.5 237.3
50...100
5.9987 7.3313 229.1
DP → dwe-point:
DP → absolute humidity:
5.5.1. Calculated values
Based on the measurement of a temperature and relative
humidity, the P18D transducer calculates dew-point and absolute humi-
dity of the following dependencies.
5.5.2. Selection of controlling values for analog outputs
The P18(D) transducers in versions P18(D)-1XX, P18(D)-
2XX, P18(D)-4XX, P18(D)-5XX are equipped with two programmable
analog outputs (voltage or current). By default, the first analog output
is set for a presentation of the ambient temperature and the second
analog output is controlled by the value of the relative humidity. The ana-
log outputs can respond to any of the measuring or calculated values