6. Fuse replacement
The milliohm- and current measurement ranges are each protected
by a separate fuse (FF500mA/250V). If one of these fuses is melted
and it is no longer possible to measure any values, the fuse must
be replaced. Use only fuses with the same technical specifications.
Proceed as follows:
1. Turn the meter off and remove the test-leads.
2. Remove the four screws on the battery compartment.
3. Remove the battery cover.
4. Remove the fuse from the fuse holder and replace it with a new
one.
5. Close the battery compartment and tighten the four screws
-46-