PressureMAT User Guide
Revision 8
Page 29/48
Time Alarm- This value is set to a default of 0000 hours. This can be set to trigger an alarm
in a set period of time and most likely will remain at 0000 hours which indicates it is not
active. Press the SELECT/PROG with Alarm Valid selected.
Use the RIGHT ARROW button on the keypad until you get to the digit that you want to
change. Use the DOWN ARROW/UP ARROW buttons on the keypad to change the values.
To change a digit from a numerical value to zero, change the value to a zero. When the correct
value has been selected, press the SELECT/PROG button and the value will be confirmed. To
escape from this menu without changing the value, press the BACK button and the main
channel menu will appear.