EasyManua.ls Logo

Pendulum CNT-91 - Period A, B Single; Ratio A;B, B;A, C;A, C;B; Time Interval a to B, B to A, a to A, B to B; Pulse Width A, B

Default Icon
189 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RF generator Pulse generator Counter
Frequency Power
Pulse
width
PRI
Start
delay
Measu-
rement
time
Sensi-
tivity
Manual
acq
frequency
F1 (GHz) F2 (GHz) Frequency
measured
+F1-F2
Tolerance P/F
1 GHz -10
dBm
100
ns
1 ms 10 ns 50 ns HIGH 1 GHz ±0.15 MHz
1 GHz -10
dBm
500
ns
1 ms 50 ns 410 ns HIGH 1 GHz ±15 kHz
1 GHz -10
dBm
1 µs 1 ms 100
ns
800 ns HIGH 1 GHz
±7.5 kHz
1 GHz -10
dBm
10 µs 1 ms 1 µs 8 µs HIGH 1 GHz
±0.8 kHz
Performance Check
7-18
USER MANUAL ● CNT 9x Series ● Rev.22 February 2020
Table 7-16. Frequency accuracy with short-pulse compensation:
Table 7-17. Pulse width measurement
RF generator Pulse generator Counter
Frequency Power
Pulse
width
Pulse
repetition
interval
Sensi-
tivity
Manual acq
frequency
Pulse width
measured
Tolerance P/F
1 GHz -10 dBm 50 ns 1 ms HIGH 1 GHz ±12 ns
1 GHz -10 dBm 100 ns 1 ms HIGH 1 GHz ±12 ns
1 GHz -10 dBm 1 µs 1 ms HIGH 1 GHz
±12 ns
1 GHz -10 dBm 10 µs 1 ms HIGH 1 GHz
±12 ns
1 GHz -10 dBm 100 µs 1 ms HIGH 1 GHz
±12 ns
Pulse width measurements
Set up Pulsed RF Pulse Width measurement on the CNT-90XL
Meas Func => Pulsed RF
=> Width => Pos => C
The following measurements should be made, by changing the required parameters.

Table of Contents

Other manuals for Pendulum CNT-91