HAFFMANS ICC
ICC Page 15 of 49
4.6 Settings
The ICC-2000 is supplied with a number of standard settings that can be adjusted by the user:
The constants used in the CO
2
formula can be refined by the user for specific beverages
and/or correlation with other reference standards.
The desired values for pressure, temperature and CO
2
content.
Values at which the instrument is to be calibrated can be indicated.
4.7 Reading and operating
Display
The ICC-2000 has a TFT display and three keys (“↑”, “↓” and “START”). The display is activated
as soon as one of the keys is used. After the last operation, the screen will remain switched on
for a variable programmable length of time.
Attention!
Wait until the values stop increasing before reading. If the display
is flashing, the value shown is unstable and still changing.
The display indicates:
Software version number immediately when switched on (“< V 1.00 >”),
Values measured for pressure (”P”) and temperature (”T”),
Value calculated for the CO
2
content (”CO2”),
Whether a value is outside the measuring range (”HHH” or ”LLL”),
When a battery needs charging (”BATTERY”),
Number under which the measurement is stored
Whether the value measured of pressure or temperature is stable (stops flashing)
Operating levels
The ICC-2000 has 4 levels can be set to the various operating levels. The four operating levels
are not secured.
Measuring screen For reading the values measured and calculated.
Configuration To set values, date/time/display, formula parameters and
language.
Calibration To calibrate pressure and check the temperature and pressure.
Logbook To view stored results and calibrations.
See the ”Software-Summary” in Appendix VII.