– 19 –
1-2-3-1. Items required
Endoscope component
• Inlet seal (OF-B190)
Reprocessing accessory
• PVE soaking cap (OE-C9)
• Cleaning brush (CS6002SN)
• Cleaning brush (CS6015ST) for EB-1570K, EB-1970K, EB-1970TK,
EB-1575K and EB-1975K
• Cleaning brush (CS3010S) (EB-1170K and EB-1990i only)
• Cleaning brush (CS-C3S)
• Cleaning adapter (OF-B155)
Other Equipment
• Protective equipment such as gloves, gowns, face masks, etc., to minimize the
risk of cross contamination.
• Detergent solution, Endozime (Ruhof Corporation)
• Clean potable water
• Basin sufcient in size to immerse the entire endoscope (at least 50 cm in width
x 40 cm in depth x 15 cm in height)
• Lint-free gauze
• 30 mL luer slip syringe
1-2-3-2. Preparation
1) Wear personal protective equipment.
2) Attach PVE soaking cap (OE-C9) to the endoscope
3) Detach Ventilation Cap (OF-C5) from the endoscope.
(1)
(2)
(1) OE-C9
(2) OF-C5
Figure 1.8
4) Fill a basin with a sufcient volume of detergent solution to completely immerse the
endoscope. Prepare the detergent in accordance with the manufacturer’s instructions
(temperature, concentration). In the case of ENDOZIME, add 30 mL of ENDOZIME
concentrate to 3.8 L (1 gallon) of clean potable water at 20°C~30°C (68°F~86°F).
Endoscope
Cleaning
1