– 38 –
1-2-5-1-3. Wrapping
WARNING
Contact the manufacturer regarding directions for use of instrument
wrap and tape.
Scope components may be wrapped and sterilized together with
the scope. However, do not attach any components other than the
ventilation cap to the scope
Secure the wrapping with a length of tape that is sufcient to prevent the
wrapping from peeling away from package. Failure to do so can result in
removal of the tape and ineffective sterilization.
1) Wrap the tray containing the scope.
2) Apply the tape lengthwise across the three sides of the wrapped container.
1-2-5-1-4. Sterilization Parameter Selection
WARNING
Contact Johnson & Johnson regarding directions for use of the STERRAD
®
NX™ system
1) Set the tray containing the scope into the sterilization chamber.
STERRAD® NX™ :
Select "Advanced" cycle and operate the sterilizer according to the instructions supplied
with the STERRAD
®
NX™.