– 31 –
Filling the suction channel with disinfecting solution
2) Attach a syringe lled with disinfecting solution to the suction nipple, and inject 25
mL of disinfecting solution into the suction channel.
(1)
(1) Suction Nipple
Figure 1.23
3) Conrm that disinfecting solution ows out from the suction channel opening on the
distal end.
4) Insert a syringe lled with disinfecting solution into inlet seal (OF-B190), and inject
25 mL of disinfecting solution into the suction channel.
(1)
(1) OF-B190
Figure 1.24
5) After injecting disinfecting solution into the suction channel, detach inlet seal (OF-
B190), and cleaning adapter (OF-B155) from the endoscope, and leave them with
the endoscope in the disinfecting solution. Open the cap of inlet seal (OF-B190).
(1)
(2)
(1) OF-B190
(2) OF-B155
Figure 1.25
Endoscope
Disinfection