– 31 –
Filling the suction channel with disinfecting solution
2) Attach a syringe lled with disinfecting solution to the suction nipple, and inject 25
mL of disinfecting solution into the suction channel.
(1)
(1) Suction Nipple
Figure 1.23
3) Conrm that disinfecting solution ows out from the suction channel opening on the
distal end.
4) Insert a syringe lled with disinfecting solution into inlet seal (OF-B190), and inject
25 mL of disinfecting solution into the suction channel.
(1)
(1) OF-B190
Figure 1.24
5) After injecting disinfecting solution into the suction channel, detach inlet seal (OF-
B190), and cleaning adapter (OF-B155) from the endoscope, and leave them with
the endoscope in the disinfecting solution. Open the cap of inlet seal (OF-B190).
(1)
(1) OF-B190
(2) OF-B155
Figure 1.25
Endoscope
Disinfection
1