– 43 –
Third rinse
7) Fill a basin with clean water and repeat steps 2-5 to perform a third rinse.
Forth rinse
8) Fill a basin with clean water and repeat steps 2-5 to perform a fourth rinse.
9) Remove all components and accessories from the water.
Purging of water from lumens
10) Using a syringe lled with air, ush the lumen of the suction valve (OF-B179) to
purge residual water. (see Figure 1.33, p41)
1-3-1-5. Drying
1) Wipe all surfaces of components and accessories gently with a lint-free gauze.
Accessory