- 12 -
10) Select Level-1 by tapping the level select button (Level 1 to 5) and adjust the air flow by tapping the
adjusting buttons so that the measuring value attains to the defined value shown in Table-2 below.
Adjust all of the levels with the same methods.
Note: 1) The selected level button turns in yellow.
2) It is not necessary to turn on the pump. When the level is selected with the level select button and the
value is changed with the adjusting button, the pump turns on automatically.
3) Just after selecting another level, the selected level has not been actuated. Its value is changed, then
the selected level is actuated.
Level Air flow(L/min)
Tolerance(L/min)
1 2.50
2 3.10
3 3.70
4 4.30
5 5.40
+/-0.10
11) Detach the air flow meter and attach the air pressure meter to the EPK-i (Refer to Photo-4).
12) Select Level-1 and record the adjusted value and the measured air pressure value.
Note: It would be about 50KPa.
13) Record all of the levels with the same methods.
14) Tap Home button to exit Processor Adjust Menu.
Photo-10 Adjusted values
Record these values
Table-2 Air flow rate
Photo-8 Processor Adjust Menu
Photo-9 Pump Level Adjustment
Level select buttons
Adjusting buttons
Pump Level button