F3 Corpus
Troubleshooting
Event
Possible cause Remedy
The wheelchair stops while being driven. The cable connection to the control
panel has come loose.
Insert the cable in the control panel.
Main circuit breaker tripped.
See page 246.
The wheelchair can only be driven at
reduced speed.
Seat lift or seat tilt raised too high.
Applies only to a powered seat lift and
seat tilt.
Lower the seat lift or seat tilt.
The wheelchair cannot be charged. Main circuit breaker switched to off
position after e.g. battery replacement.
See page 246.
The charging circuit breaker has tripped. Wait five minutes, the circuit breaker will
automatically reset.
8.2 Diagnostics R-Net LCD control
panel
8.2.1 General
When an error or a fault occurs in the wheelchair’s electronics,
information about it is displayed in the control panel display. This
information can then be used to diagnose where the error, or fault,
occurred and its cause.