EasyManuals Logo

Pfeuffer HOH-EXPRESS HE 50 User Manual

Pfeuffer HOH-EXPRESS HE 50
26 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
Operation
16 | 26 Operating instructions HE 50 | Revision 7
After approx. 7 seconds, the result is shown with the first two letters of the product:
Wh 14.5%
Turning the dial switch causes another measurement to be
performed immediately thereafter.
Pressing the measurement key several times can produce a slightly
different result, because the calibration is configured for an individual
measurement.
Change the ratchet position to open the measurement cell.
Hold the measurement cell at the side and unscrew the top part by making short
movements with the ratchet.
NOTICE
Remove moist milled material from the measurement cell
immediately after performing the measurement!
Clean the measurement cell and the spindle after each measurement,
see chapter 6.1.
The measurement cell is equipped with special fluting to allow all
grain-type products to be crushed.
To avoid damage, the measurement cell is not allowed to be screwed
together bringing the milling disks into contact without any material
fill. Place the plastic disk supplied with the product around the spindle
as protection.
Figure 5: Protection against touching the milling disks
1
2
3
Item
Designation
1
Measurement cell top part
2
Plastic disk
3
Measurement cell bottom part

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pfeuffer HOH-EXPRESS HE 50 and is the answer not in the manual?

Pfeuffer HOH-EXPRESS HE 50 Specifications

General IconGeneral
BrandPfeuffer
ModelHOH-EXPRESS HE 50
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals