17
F05: <Function> Compressor is turned on with forcibly delayed (by minute increment)
when the power retrieves from the power failure.
<Operation> Input “F05” and press SET key to display “003” (initial).
Change the value in the range “003”~”015” with key and key.
Press SET key to store the value and revert to chamber temperature
display.
F06: <Function> Service code input and initialization of accumulation time
<Operation> Input F06 and press SET key to display “000” (initial value).
Set to “384” with key and key. Press SET key to store the
value and revert to chamber temperature display.
<Cancel> Input F06 and press SET key to display “384”.
Change to “000” with key and key. Press SET key to store
the value and revert to chamber temperature display.
Service code is cancelled when the power is ON and OFF, or
accumulation time is reset.
F07: <Function> To match temperature of temp. sensor with temperature of center at
the chamber.
<Operation> Input service code in F06 prior to use this mode.
Input F07 and press SET key to display “00.0” (initial value).
Change to the desired value with key and key in the range of
“-4.9”~”04.9”. Press SET key to store the value and revert to
chamber temperature display.
F12: <Function> To display the temperature of temp. sensor
<Operation> Input service code in F06 prior to use this mode.
Input F12 and press SET key to display alternately F12 and “XX.X”
(chamber temperature). Press SET key to revert to chamber
temperature display. 3 digits indication. Minus “-“ is not displayed.
Ex) “-79.5㷄” 㸢㩷 indicated as “79.5”
F14: <Function> To display the temperature in comp. sensor
<Operation> Input service code in F06 prior to use this mode.
Input F14 and press SET key to display alternately F14 and “XX.X”
(present temperature in comp. sensor). Press SET key to revert to
chamber temperature display.
F17: <Function> Non-volatile memory initialization, model code change
<Operation> Service code should be input in F06 prior to use this mode.
Input F17 and press SET key to display “001”. Change the value
with key and key. Press SET key to store and revert to
chamber temperature display.
Model code ‘003’: MDF-C8V1
<Initial values in non-volatile memory>
Zero adjustment for temp. sensor
Compressor forcibly OFF
Ring back time
Setting temperature
High temp. alarm setting temp.
Low temp. alarm setting temp.
Linkage between remote alarm
and buzzer
Compressor delay time
Communication ID
Communication mode
Temp. alarm delay time
Key lock
: 0㷄
: 8 min.
: 30 min.
: -80㷄
: +10㷄
: -10㷄
: OFF
: 3 min.
: 000
: 000
: 15 min.
: OFF