18
F21: <Function> Serial communication ID setting
<Operation> Input F21 and press SET key to display “000” (initial).
Change the value in the range “001”~”255” with key and key.
F22: <Function> Serial communication mode setting
<Operation> input F22 and press SET key to display “000” (initial value)
Change the value with key and key.
Press SET key to store the value and revert to chamber temperature
display.
Control mode (the 3
rd
digit)
0: Local (initial)
1: Remote
Baud rate (the 2
nd
digit)
0: 2400bps (initial)
1: 4800bps
2: 9600bps
Note) You cannot be changed SV if control mode is set in “Remote”.
F24: <Function> Linkage between remote alarm and buzzer
<Operation> Input F24 and Press SET key to display “000” (initial).
Change to the desired value with key and key.
Press SET key to store the value and revert to chamber temperature
display.
000: Remote alarm is outputted simultaneously with buzzer
001: Remote alarm is not outputted simultaneously with buzzer
F25: <Function> Ring back time setting
<Operation> Input F25 and press SET key to display “030” (initial).
Change to the desired value with key and key.
Press SET key to store the value and revert to chamber temperature
display.
000: Auto recovery OFF
010: Recovers after 10min. passed
020: Recovers after 20min. passed
030: Recovers after 30min. passed
(initial)
040: Recovers after 40min. passed
050: Recovers after 50min. passed
060: Recovers after 60min. passed
F26: <Function> Actual running rate is displayed (unit: %)
<Operation> Service code should be input in F06 prior to use this mode.
Input F26 and press SET key to display alternately F26 with “XXX”
(present actual running rate).
Press SET key to revert to chamber temperature display.