EasyManua.ls Logo

Philips CARDIOMD - Page 84

Philips CARDIOMD
466 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Detector Calibration
4-28 CardioMD Service Manual 1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
15. Click Exit to close the Uniformity Calibra-
tion page.
16. Replace the FOV mask with the collima-
tor.
17. Slide the collimator on the other detector
back in place.
18. Press the hand controller Collision Over-
ride button and wait for a minimum of 4
seconds.
19. Repeat the procedure outlined above to
calibrate the other detector.
20. Click Exit to close the Uniformity Calibra-
tion page.
Extrinsic Uniformity Calibration Procedure
Procedure Details
1. Ensure that the CardioMD system is in the
patient load position and that the acquisi-
tion PC is displaying the Persistence page.
2. Click the Tools button.
The Calibration & Configuration menu ap-
pears.
3. In the Calibration & Configuration menu,
click Uniformity calibration.
The Uniformity Calibration page appears.
4. Specify the energy settings for the uni-
formity correction table to be acquired.
5. Select Extrinsic.
6. Select the detector to be calibrated.
The uniformity correction table is assigned a
filename of the form:
U.coll.ew1.ew2.ew3.detectorNum
Where:
coll is set to LEGP, LEHR or NONE, depending
on the collimator mounted on the detector.

Table of Contents