EasyManua.ls Logo

Philips CARDIOMD - Page 85

Philips CARDIOMD
466 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration
1MAN0158-H11 04-2005 CSIP Level 1 CardioMD Service Manual 4-29
Procedure Details
Ew1, ew2, ew3 are set to the values entered for
the isotope peaks.
DetectorNum is set to either 1 or 2, depend-
ing on which of the detectors the table is ac-
quired for.
7. Click Continue.
8. When prompted, click OK to accept auto-
mated motions.
The detectors are automatically positioned for
extrinsic uniformity calibration of the selected
detector.
9. Place a sheet source or a flood phantom
filled with the appropriate isotope and uni-
formly mixed on the surface of the colli-
mator on the detector.
10. Verify that the count rate is less than 40
kc/s.
Note. When the CardioMD AC option is
mounted on the system, it is not possible to
use this calibration position. See Chapter 7
CardioMD AC Option of the CardioMD Op-
erator’s Manual for further information.
11. Click Start to start the uniformity calibra-
tion.
When the calibration is completed, the table is
automatically stored.
12. Click Exit to close the Uniformity Calibra-
tion page.
13. Repeat this procedure to calibrate the other
detector.
Detector angle = 135°
2
1
Place flood phantom or sheet
source here for detector 1
Place flood phantom or sheet
source here for detector 2

Table of Contents