IMMUNITY test IEC 60601 test
level
Compliance
level
Electromagnetic
environment – guidance
Electrical fast
transient/burst IEC
61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output Lines
± 2 kV for power
supply lines
± 1 kV for input/
output Lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge IEC 61000-
4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
Mains power quality
should be that of a typical
commercial or hospital
environment.