Declaration – electromagnetic immunity
Guidance and manufacturer’s declaration – electromagnetic immunity.
The device is intended for use in the electromagnetic environment specied below.
The customer or the user of the device should assure that it is used in such an environment.
IMMUNITY test IEC 60601 test
level
Compliance
level
Electromagnetic
environment – guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile.
If oors are covered with
synthetic material, the
relative humidity should be
at least 30 %.