IMMUNITY test IEC 60601 test
level
Compliance
level
Electromagnetic
environment – guidance
Power frequency
(50/60 Hz)
magnetic eld IEC
61000-4-8
3 A/m 3A/m Power frequency magnetic
elds should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Note: UT is the a.c. mains voltage prior to application of the test level.