Philips EP Series
Sequence of
actions by user
Reaction of the appliance
PASS FAIL Cause of failure Solution
EV
Press and release
K to toggle the
EV.
If it was closed,
then will be
open.
• L White ON in UI panel +
• BoxK White in Display +
BoxT Green in Display +
BoxT Green in Display
• It’s possible to hear the “click”.
The “click” is no
audible. The EV
remain closed
Wiring of the EV not
connected
Check the wiring
Wiring of the EV
damaged (open)
Change Main
Board
EV damaged Change the EV
The drive of the EV in
the Main is damaged
Change Main
Board
Press and release
K to toggle the
EV.
If it was open,
then will be
closed.
• L OFF in UI panel +
• BoxK White in Display +
BoxT Green in Display +
BoxT Green in Display
• It’s possible to hear the “click”.
The “click” is no
audible. The EV
remain open
EV damaged Change the EV
Test mode