SERVICE & TECHNICAL INFORMATION 1039055, VER. 06
SYMPTOM CAUSE VERIFICATION
CORRECTIVE
ACTION
•
Fluctuations in
oxygen pressure
•
Fluctuation in flow
ball
Pressure Regulator
failure
Valve Solenoid failure
•
Verify the flow from
regulator is correct
•
Perform Outlet /
Regulated Pressure
Test
•
Verify the Valve
Solenoid is working
properly
•
Verify the valve
wiring harness is
connected at the
Main PCA
•
Replace the
regulator
•
Replace the Valve
Body
•
Reconnect the
wiring harnesses
and retest
•
If problem
continues, install
old valve body
and new pilot
l
•
Unit will not turn on
No power to the unit
•
Verify the power cord
is connected to the
wall outlet
•
Verify there is power
in the wall outlet
•
Verify if the outlet is
connected to a light
switch and the switch
is in the ON position
•
Connect the
power cord to the
wall outlet
•
Check the
household fuse or
circuit breaker
•
Move the power
switch to the ON
position