SERVICE & TECHNICAL INFORMATION 1039055, VER. 06
7.4 SYSTEM PRESSURE TEST TABLE
SYMPTOM CAUSE VERIFICATION
CORRECTIVE
ACTION
System Pressure Test
•
Abnormal reading
•
High-pressure
reading above 25
psig @ 60 Hz @ 5
LPM flow.
•
Check oxygen
percentage at 5
LPM. If Low?
•
Check for proper
valve cycling
•
Replace Sieve
Canister
•
Fluctuating
pressure
•
Difference in Peak
pressure is > 1 psig.
•
Check oxygen
percentage at 5
LPM. If Low?
•
Check for leaks
•
Replace Sieve
Canister
•
Low pressure
reading
•
Low pressure reading
below 20 psig @ 5
LPM
•
Check oxygen
percentage at 5
LPM. If Low?
•
Check for leaks
•
Check tubing
•
Check fittings
•
Check for dirty filters
•
Check Compressor
for System Pressure
•
Replace tubing
•
Replace fittings
•
Replace filters
•
Replace
Compressor